Quick Installation Guide
00825-0100-4697, Rev DB
September 2011
Rosemount 848T
7
Option 2:
1. Connect sensor wiring shield(s) to the transmitter enclosure (only if the enclosure is
grounded).
2. Ensure the sensor shield(s) is electrically isolated from surrounding fixtures that may be
grounded.
3. Ground F
OUNDATION fieldbus signal wiring shield at the power supply end.
Grounded Thermocouple Inputs
1. Ground sensor wiring shield(s) at the sensor.
2. Ensure that the sensor wiring and F
OUNDATION fieldbus signal wiring shields are
electrically isolated from the transmitter enclosure.
3. Do not connect the F
OUNDATION fieldbus signal wiring shield to the sensor wiring
shield(s).
4. Ground F
OUNDATION fieldbus signal wiring shield at the power supply end.
Analog Device Inputs
1. Ground analog signal wire at the power supply of the analog devices.
2. Ensure that the analog signal wire and the F
OUNDATION fieldbus signal wire shields are
electrically isolated from the transmitter enclosure.
3. Do not connect the analog signal wire shield to the F
OUNDATION fieldbus signal wire
shield.
4. Ground F
OUNDATION fieldbus signal wire shield at the power supply end.
Transmitter Enclosure (optional)
Ground in accordance with local electrical requirements.
Sensor Wires
Power
Supply
Shield ground points
848T
Sensor Wires
Power
Supply
Shield ground points
848T
Power
Supply
848T
Analog
Device
Analog Device
Power Supply
4-20 mA loop
FOUNDATION
fieldbus bus
Shield ground points